2인용 글로브 박스

Model: MR-Lab2000-2400×7

Specifications:

StructureIntegrated structure (Purification System integrated under the box)
Power Supply110 - 220V AC, 50/60Hz, Single-phase
Glove box BoxBox Material: 304 Stainless Steel (Type 304), thickness 3mm
Box Size(excluded antechamber and support): 2400mm(L) x 750mm(W) x 900mm(H) 94.49 " x 29.5" x 35.4"
Box Support height: 900mm(35.43" )
TrayStandard Layout : Two layers in box.

Option/Add(tool boxes or other) : Various tool boxes, can be placed according to user needs.
Front windowTilt design operating window, removable safety toughened glass front window, 8mm thickness,Wear resistance, corrosion resistance, good light transmission, sealing ring using 3/8 inch thick OMEGA sealing ring.

Option (Window Open) : The front window can be designed to open instead of fix, in order to put in a variety of instruments (custom function).
Environment of Box insideH2O <1 ppm (20°C/68 ℉, 1 atm)
O2 <1 ppm (20°C/68 ℉, 1 atm)
AntechamberControl way: manual
Design: two circular antechambers
Main chamber size: Φ360mm (Dia.)x 600mm(length)(location: right-side of box)14.17 "(Dia.) x 23.62" (L)
Mini chamber size: Φ150 mm(Dia.) x 300mm(L)(location: right-side of box) 5.9 "(Dia.) x11.8" (L)
Vacuum degree:<=-0.1MPa(right side antechambers included main chamber and mini chamber, all of that are need vacuum and display. The vacuum pressure gauge brand is German WIKA。

Option/add auto control way: It will automatically control the antechamber by time setting and the number of times for gas exhaust and filling.

Option/Add: heating mode :
Max temperature:200℃/ 392℉
Control accuracy:±2℃/±3.6℉
The main chamber and mini chamber are both can adding the heating function (compared with standard antechamber the size is a little different)
Working gas and regeneration gasWorking gas: 99.999% high pure N2, Ar, He and other inert gas
Control gas: 99.999% high pure inert gas
Regeneration gas: mixing gas of working gas and H2,hydrogen(H2)Volume accounted for 5-10%
If the pure system only has moisture removal function, the regeneration gas is the same with working gas
Gas purification systemSingle pure column:
German BASF Oxygen removal material, US UOP high moisture absorb material. The regeneration process of purification system is auto control and automatic moisture and oxygen removal function. Gas purity can be sustained and keep long-term:H2O<1ppm, O2<1ppm。

Option: dual columns purification system (one in use and the other is reserve)
Why choose duel pure column? 1、The Pure column need to be regenerated after saturated by absorbing the oxygen and moisture. The regeneration time is 16hours. If there is no reserved column, it may stop working. (we recommend dual column purification system to factory or other unit which need glove box working Continuously); 2、The pure capacity of dual pure columns is two times as much as single column
Pressure control systemBox pressure can be auto controlled by PLC touch screen
Control accuracy:±1Pa
Can also be manually controlled by foot switch
Filter systemThe ends of inlet and outlet both are provided with filter
The filter precision is: 0.3 mu m
Closed circulating fanHigh speed frequency conversion fan, flow 0-100m3/h, microprocessor control, Stainless steel device for pressure testing.
With water cooling device (can equipped with water cooler and external refrigeration equipment)
Control system SIEMENS colors touch screen (6 inches), PLC control system, Bilingual in Chinese and English can be switched
Pressure sensor: -2500~2500Pa touch screen, accuracy±1Pa
Pressure control: control the box pressure to one setting point (scope can be set)
measurements Moisture analyzer: US AVCRAY brand, 0~500ppm touch screen display, precision 0.1ppm

Oxygen analyzer: US AVCRAY brand, 0~1000ppm touch screen display, precision 0.1ppm
Vacuum Pump Britain Brand EDWARDS, pumping speed 8.4-12m3/h, Model: RV12
Extreme pressure 2 x 10-3 mbar/1.5 x 10-3 Torr
Inlet flange NW25
Outlet flange NW25
Noise level: 50 Hz 48 dBA
Glove Brand: GUARDIAN
Material: Butyl
illumination systemEquipped with Fluorescent lamp. Brand: PHILIPS
Solvent removal systemEfficient adsorption of activated carbon, can effectively remove trace organic solvents;
Spare KF40 standard jointSpecial design to KF40 Joint, you can lead the water into the box
Can be connected to other devices (low temperature refrigerator, film coating machine, microscope, solvent purification system, low temperature cold trap, heating furnace, etc.) low temperature refrigerator

microscope

Solvent Purification System

Coating machine:
Inside connection(one door of Coating machine is inside of glove box)
Outside connection
Equipment size and room requirementsOverall size :
3050mm(L) x 1000mm(W) x 1950mm(H)
120.08" x39.37" x 76.77"

Room size:
2200mm(L) x 1500mm(W) x 2000mm(H)
133.86" x59" x 78.74"
Wooden Package size and weight3385mm(L) x 1145mm(W) x 2300mm(H)
133.27" x 45" x 91" (5.8M3)

1020 kg (2197 lbs)
NoticeWhen the gas is circulating, must ensure that the piping of the circulation system is clear and cannot be blocked (if the color of the filter is changed from white to black, the filter should be replaced timely).
No corrosive gas in the glove box.
It needs to regenerate the column when the moisture and oxygen indicators cannot meet the work requirements.
If the corrosive gas is released from inside of glove box.(such as LiPF6 electrolyte), please do not use. you must put the bottle filled with liquid seal.
Vacuum pump needs to be checked the oil level and maintain in regular time.

Main Configuration (1 Unit)

No.ProductQuantityBrandNotes
1Box2 PCSMRBESTSS 304, Thickness 3mm
2Ante Chamber1 setMRBESTSS 304
3Control System1 setGermany SIMENS Omron Electronics etc.Imported then do assembly
4Circulation System (With purification and regeneration system)1 setGermany BASF, American UOP Purification material, German Boyd ValvesImported then do assembly
5Display System1 setGermany SIMENSImported
6Gloves2 pairUSAImported
7Pipe Interface1 setFrance LegrisImported
8Pressure Gauge1 setGermany WIKAImported
9PLC Moisture Analyzer1 PCUSA AVCRAYImported
10PLC Oxygen Analyzer1PCUS AVCRAYImported
11Vacuum Pump1 PCBritain RV12Imported
12Organic Solvent Adsorption1 setMRBESTActivated Carbon + Activated Alumina